Performance characteristics of five automated CA 19-9 assays

Sonia L. La'ulu, William L. Roberts

Research output: Contribution to journalArticlepeer-review

43 Scopus citations

Abstract

Serum concentrations of cancer antigen (CA) 19-9 can be useful in monitoring response to therapy in pancreatic cancer. The objective of this study was to evaluate 5 automated CA 19-9 assays: ARCHITECT i2000 (Abbott Diagnostics, Abbott Park, IL), ADVIA Centaur (Bayer Diagnostics, Tarrytown, NY), UniCel DxI 800 (Beckman Coulter, Fullerton, CA), IMMULITE 2000 (Diagnostic Products, Los Angeles, CA), and Elecsys E170 (Roche Diagnostics, Indianapolis, IN). All methods were evaluated for limit of detection, linearity, imprecision, method comparison, and reference intervals. Limit of detection results were all below 2.0 kU/L and met the manufacturers' claims. Linearity had deviation from target values that ranged from 4.5% to 26.7%. All methods showed acceptable imprecision with total coefficients of variation less than 8%. Method comparison by Passing-Bablok analysis resulted in slopes ranging from 1.00 to 2.06 and correlation coefficients of 0.85 to 0.98. Between 97.6% and 99.2% of results from healthy volunteers were less than 35 kU/L. All methods show acceptable analytic performance. © American Society for Clinical Pathology.
Original languageEnglish (US)
Pages (from-to)436-440
Number of pages5
JournalAmerican Journal of Clinical Pathology
Volume127
Issue number3
DOIs
StatePublished - Mar 1 2007
Externally publishedYes

Bibliographical note

Generated from Scopus record by KAUST IRTS on 2023-09-20

ASJC Scopus subject areas

  • Pathology and Forensic Medicine

Fingerprint

Dive into the research topics of 'Performance characteristics of five automated CA 19-9 assays'. Together they form a unique fingerprint.

Cite this